GoTo HomePageHarwood Engineering Company, Inc.

The Piston Gage as a Precise Pressure-Measuring Instrument

by D.P. Johnson and D.H. Newhall

The errors which must be considered for accurate measurements of pressure with piston gages are discussed. A new design of piston gage is described which permits the operator to control the clearance between piston and cylinder at any operating pressure. Instrumental errors are thus substantially reduced, particularly those resulting form elastic distortion. A check list of errors inherent to piston gages is presented. These errors are to be considered when instruments are used to accuracies better than ½ of 1 percent. The need for a check on the longtime variations of the piston gage is supplied by the fixed points, changes of state of some very pure materials which take place at definite pressures. Such are the melting pressure of solid mercury at 0°C, the melting pressure of water at 30°C, the transition between crystalline states of bismuth. As an indication of its possibilities the experimental model of the controlled-clearance piston gage was used to measure the melting point of mercury at 0°C. Results were satisfactory agreement with Bridgman's value.

Introduction

Industrial engineers sometimes ask how they may benefit by the concern of the National Bureau of Standards over a primary standard of pressure with an accuracy of 0.01 percent or better. An industrial engineer may be concerned with processes which require the control of pressures within, say, ± 1 percent. To be sure of holding the pressures within the required range, he will have on his control panel, pressure gages with an accuracy of, say, 0.3 percent. The manufacturer of the pressure gages will have in his production line a calibration bench. Here the gages will be checked against standards which should be better than the instruments calibrated by at least a factor of 3, preferably a factor of 10, that is, from 0.1 to 0.03 percent. The manufacturer should have in his standards room instruments somewhat better than the bench standards, with an accuracy better than 0.3 percent, preferably about 0.01 percent.

There are many research laboratories in industry engaged in getting fundamental engineering data with an accuracy of 0.1 percent or better. Phenomena which are sensitive to pressure may require instrumentation accurate to 0.03 or even 0.01 percent.

The National Bureau of Standards should have primary standards which are somewhat more accurate than any instrument portable enough to be brought in for calibration. There are pressure-measuring instruments which are capable of an accuracy of 0.01 percent at pressures up to a few thousand pounds per square inch. The primary pressure standards against which these are calibrated should be definitely better, with an accuracy of 0.003 percent. The possible accuracy drops off as the pressures increase, and considerable effort is required to develop primary  standards adequate to the present requirements in the high-pressure field.

Pressure Standards Apparatus

In a primary standard the measurement of any physical quantity is referred directly to the standards of mass, length, and time. The device should be so simple in principle that all of the errors inherent in the instrument either can be eliminated or evaluated. Two commonly used primary standards are the mercury column and the dead-weight piston gage. The choice of instrument is not limited by accuracy, but by convenience. Twenty pounds per square inch is approximately equal to the head of a convenient 40-inch column of mercury. To measure 30,000 psi would require a column of mercury about 1 mile high.

The first extensive use of the free-piston gage was by Amagat (1).³ The basic form and two modifications are in current use. They are illustrated in Figure 1, Figure 2, and Figure 3.

Simple Piston Gage
The most familiar form of piston gage is shown in Figure 1. It is a fitted piston-and-cylinder arrangement with means to apply known weights on the piston. The piston is rotated or oscillated to reduce friction. When the pressure applied to the piston develops a force sufficient to support the dead weights, they will be lifted. The weights may be placed on top of the piston or be suspended below it by a yoke. The weights may rotate with the piston or the piston may be rotated or oscillated independently of the weights. It is necessary for some fluid to leak past the piston for lubrication. The leak should be kept small to provide sufficient floating time to establish equilibrium and to make the required observations. This is particularly important at high pressure.
Differential Piston
Professor Michels (2) of Holland has designed a sensitive piston gage for high pressure, Figure 2. It is a series of two differential-area pistons both carrying dead weights. The first piston (A) is suitable to pressures of 3000 kg/cm² (43,000 psi).

When its output is piped to the upper side of the second piston (B), the force developed is additive to the dead-weight load. The second piston makes the unit suitable to 10,000 kg/cm² (142,000 psi). Leakage is controlled by an interface fit between piston and cylinder. At high pressure the piston becomes free to rotate. The use of similar pistons of various sizes permits measurements over a wide range.

Re-entrant Cylinder
Professor Bridgman (3), early in his work developed a free-piston gage with re-entrant cylinder, Figure 3. The bottom tip portion of the piston (p) and outside surface of the cylinder (C) are subjected to the measured pressure. The bore of the re-entrant cylinder (C) grows smaller as pressure increases, regardless of the wall ratio (W = OD/ID). Since the clearance between the piston and cylinder diminishes with increasing pressure, and the leakage is reasonably small, the upper limit of the range is determined by the initial clearance. The initial clearance was chosen so that the crevice closed completely at approximately 210,000 psi, and the instrument was used to approximately 190,000 psi to a precision of approximately 0.1 percent. The fluid used was a mixture of glucose, glycerin, and water.

The literature abounds with descriptions of instruments which are variations of the basic concept of the simple piston gage. Mostly, they are clever arrangements to allow the operator to handle small weights. It is probably fair to say that uncertainties are built into the devices in proportion to the departure from the simply loaded piston.

Controlled-Clearance Piston Gage

The clearance between the cylinder and piston must be kept small if the piston gage is to be useful on the score of (a) knowledge of the effective area, (b) high sensitivity, (c) manageable leakage of the fluid past the piston. The large elastic distortions, particularly in the cylinder, have set severe limitations on the useful range.

The controlled clearance piston gage, Figure 4, permits a size adjustment of the cylinder bore to the desired clearance. The measuring piston (a) is contained in a jacketed cylinder (b). An independent pressure is used between the jacketed and jacketing cylinder (c) to deform the jacketing cylinder to the desired clearance. The pressure in the jacketing cylinder is sealed off by Bridgman unsupported area rings (d) whose initial seal is accomplished by a mechanical load from nut (e).

The relation between the pressure PJ required in the jacketing cylinder and the pressure PM being measured is expressed approximately by

PJ = K + LPM
in which K is the pressure required to close the crevice at zero internal pressure. The value of L depends on the position of the ring of minimum clearance. If the ring is near the top of the cylinder, L will be about 0.7; if away from either end, the preferred condition, L will be about 0.35. Depending on the initial clearance, the jacketing pressure can be either greater or lesser than the pressure being measured. The jacketed cylinder and piston are disposed in the design so that the seal between them will be accomplished sufficiently far from the ends of the piston to avoid local distortions which result from end effects. To avoid end effect further, the jacket cylinder is "bellmouthed" at both ends of its bore.

The free-falling time is controlled readily by minor adjustments of jacket pressure. If the close clearance is maintained, the effect of the distortion of the cylinder is eliminated. It will be shown later that this is a notable improvement since in other designs the elastic distortion of the cylinder is one order greater than the distortion of the piston. A still further improvement probably will result from the use of carboloy pistons. Otherwise the apparatus is more elaborate than unique. The elaborations take advantage of well-known design points with a view to maintaining accuracy and precision, and to securing ease in manipulation.

The dead-weight system, Figure 5, is patterned after the 10,100-lb. dead-weight system in the Engineering Mechanics Section at the National Bureau of Standards (4). The dead weights themselves (f) are made of austenitic stainless steel for long-time stability and freedom from oxidation, corrosion, and magnetism. The loading system has 1000 lb. of dead weight, consisting of nine 100-lb. platters held in a chain suspension attached to the yoke (g). The weights are lifted on and off the suspension by the power cylinder (h) on the base. Smaller 10- and 5-lb. weights (i) are provided for hand-loading on the yoke. The piston is rotated either clockwise or counterclockwise by means of a pulley (j) and motor arrangement to eliminate the corkscrew effect. It is patterned after that described by Meyers and Jessup (5). The yoke suspension rods (k) are insulated electrically from the upper cross rail (l). Thus, by electrical sensing, there is assurance that the weights supported by the piston are free of other parts.

The measuring piston and cylinder are so arranged that they may be removed readily and interchanged with similar components for other ranges. The cost of the piston assemblies is on the order of 3 percent of the total cost of the instrument, while the piston itself represents only a fraction of 1 percent of the total cost of the instrument. The change-over from one piston size to another is done readily with hand tools.

Errors of a Piston Gage

Definitions
A pressure is a force per unit area.A force of 1 lb. will be defined as that which will accelerate a mass of 1 lb at the normal acceleration of gravity, 980.665 cm/sec² or 32.174 ft./sec². The U.S. pound is defined as the 0.4535924277 part of the international kilogram. The U.S. standard inch is defined as 100/3937 international meters (6).

To obtain the pressures as so defined, the indications of a practical dead-weight piston gage are subject to a number of corrections. The more important of these, which will be of concern to those who hope to measure a pressure to a part in a thousand, will be discussed in some detail. Some of these errors can be evaluated and corrections applied, but in every case there will remain some residual uncertainty which may or may not be important.

Effective Area
The effective area of a piston gage is the mean of the area of the cylinder and that of the piston, provided the piston is concentric with the cylinder and falling at a rate at which the volume displaced by the piston is equal to the leak between it and the cylinder. This is true for any clearance and for fluid of any viscosity. If, because of other volume changes in the system, the piston is falling at any other rate, of the same order of magnitude, the effective area will be changed by an amount of the order of the square of the clearance distance.
Elastic Distortions
The following formulas may be found with the assistance of standard texts on elasticity (7): In the piston gage the pressure varies along the length of the crevice between the piston and cylinder. At the bottom it equals the pressure being measured; at the top it falls to zero. The fall of pressure probably will be concentrated in a small portion of the length. The effective pressure in the crevice, to be used in calculating the distortion, will depend on the location of the region of falling pressure. If this region is at the bottom of the piston, the effective pressure will be small. If it is at the top, so that the crevice is filled with liquid at high pressure, then the effective pressure may be nearly as high as that being measured.

Small variations in dimensions may have a great effect on the position of the pressure drop. (It is even possible that the position may be a function of pressure.) The variations may occur because of the design, because of irregularities in machining, or because of distortions in use. For example, when a piezoelectric gage is calibrated by a sudden release of pressure, the piston of a gage connected to the system will drop violently. If it is stopped at the blind end of a cylinder it may mushroom slightly, Figure 6a. If it is stopped by contact with the open end of the cylinder, the latter may develop a burr, Figure 6b. Therefore it can be said with certainty, only that the effective pressure in the crevice is less than the pressure measured. The distortions of the piston and cylinder may be calculated on the basis of an effective pressure in the crevice which is the mean between that at the two ends, provided that the fall of pressure takes place in a region remote from the ends of the piston, and of the bore of the cylinder. "Remote" can be taken as one diameter in the case of the piston, and three bore diameters for the cylinder.

Piston
Case 1
When the pressure in the crevice is one half the pressure being measured, we have the following change DA in the area A using Equation 1, a value of 0.28 for Poisson's ratio, and 30 x 106 psi for Young's modulus:
Equation 3
At 200,000 psi this amounts to -0.11 percent.
Case 2
If the fall of pressure is at the top of the piston and occurs over a length which is small compared to the diameter, Pc=P
Equation 4
At 200,000 psi this amounts to -0.58 percent.
Case 3
If the fall of the pressure is at the bottom of the piston Pc=0
Equation 5
At 200,000 psi this amounts to +0.38 percent.
Piston Gage, Simple Cylinder, and Piston
Here the cylinder is subjected only t the pressure in the crevice. Equation 2 is applicable, with the terms Po= 0, Pe= 0, and Pa= Pc= P/2. For a wall ratio W=3, the change in area is:
Equation 6
At 200,000 psi this amounts to +1.02 percent.

The change in the effective area of the gage will be the mean of the changes in the cylinder and piston (Case 1) or

Equation 7
At 200,000 psi this amounts to +0.46 percent.

If the piston is expanded at the lower tip (Case 3), in a blind cylinder, Figure 6a, the distortion at 200,000 psi might vary from +0.19 percent at the bottom of the stroke, to +0.70 percent when the tip of the piston has risen several bore diameters away from the end of the cylinder, Figure 6c.

Piston Gage With Re-Entrant Cylinder
The end face and outside of the re-entrant cylinder are subjected to the pressure measured so that Pe= Po = P. If Pa = P/2 the distortion DA/A of the cylinder of wall ratio 3 is
Equation 8
At 200,000 psi this amounts to -0.8 percent.

The change in the effective area of the piston gage (piston Case 1) is

Equation 9
At 200,000 psi this amounts to -0.46 percent.
Controlled-Clearance Piston
In this instrument the dimensions of the cylinder change with those of the piston. The piston extends outside the section of close clearance with the cylinder by several diameters, so that the conditions for using the mean value of the pressure in the cylinder are satisfied, Figure 6d. Therefore the elastic distortion is that of the piston alone. For steel with v=0.28 and E=30 x 106 psi, the change in area is
Equation 10
At 200,000 psi this amounts to -0.11 percent.

For carboloy, with v=0.22 and E=88 x 106 psi

Equation 11
At 200,000 psi this amounts to -0.08 percent.
Measurement of Area
In the size range used in pressure gages, the piston diameters can be measured to 10 microinches or a little less. The corresponding uncertainties in the measured error are given in Table 1.
 
Table I. Effect of a 10-microinch Error in Diameter
Nominal Area, square inch Approximate Diameter, inch Error in Area, percent
1/200 0.0798 0.025
1/100 0.1128 0.018
1/50 0.1596 0.012
1/20 0.2523 0.008
1/10 0.3568 0.006
1/5 0.5046 0.004
1/2 0.7979 0.0025
The gain in accuracy by increasing the size of the piston is somewhat more than that shown in the table since more precise methods of measurement are available for the largest size.
Measurement of Mass
The weights should be of the materials and quality required for the better commercial-scale weights. The material should be nonporous, corrosion and wear resistant, nonmagnetic. Weights usually are adjusted by the manufacturer to about 0.01 percent. They can be calibrated to 0.001 percent. In precise work, these corrections should be applied.
Gravity
The readings of Bourdon-tube pressure gages, manganin-wire gages, and such phenomena as elasticities, viscosities, phase changes, and chemical reactions, are functions of the absolute pressure. They are not sensibly affected by the local value of gravity. The pressure measured by a piston gage loaded with a particular set of dead weights is proportional to the local value of gravity. The convention here used is that the readings of elastic and dead-weight pressure gages should correspond when the dead weights are acted upon by gravity at the standard value of 980.665 cm/sec².

In the southern part of the United States, the value of gravity will be more than 0.1 percent smaller than the standard value. In Canada and the northern part of the United States the local value will be higher.

If the latitude f and the elevation h in feet above sea level are known, the gravity correction Cg will be given approximately by:

Cg = - R (0.00261 cos 2f + 0.000000095 h +0.00006)
where R is the reading of the piston gage.

This correction usually will compensate for gravitational variation with an error of less than 0.005 percent in any part of the United States.

For example: at Boulder, Colorado, the reading on a piston gage, before gravity correction, was 101,200 psi. The latitude was 40°01' (sin 80°02' = +0.1730) the elevation 5350 ft.

Cg= -101,200 (0.00261 x 0.1730 + 0.000000095 x 5350 + 0.00006) =
-101,200 (0.00102) = -103
so that the corrected pressure = 101,200 - 103 = 101,097 psi.

In this case had the value observed at this location by the U.S. Coast and Geodetic Survey been used, the corrected pressure would have been 101,091 psi. This difference of 6 psi is reasonable since the errors in the formula are largest in mountainous country.

Air Buoyancy
The mass of the piston and loading weights is determined on the ordinary commercial basis, i.e. by weighing in air, on an equal-arm balance, against standard brass weights (density 8.4 grams per cc). In precise work, the mass of the load on the piston should be reduced by an amount equal to the mass of the air displaced by the weights.

The density of air at room temperature and sea-level pressure is about 0.0012 gram per cc; and the mass under these conditions will be reduced by 1 part in 7000.

Temperature
The effective area of a carbon-steel piston and cylinder may be expected to increase about 13 ppm per degree F temperature rise; for a stainless-steel piston and cylinder the increase in area will be about 18 ppm per degree F.

The temperature of the piston is somewhat higher than that of the surroundings because of the dissipation of energy in the fluid escaping between the piston and cylinder. The temperature rise in the piston is difficult to estimate because of the uncertainty in the thermal transfer between the piston and the exterior. Since the dissipated energy is proportional to the rate of leak, measurements can be made at various rates, and an extrapolation made to zero leak.

Aging
Over a period of years the dimensions of the piston and cylinder may change by as much as 1 part in 1000 because of aging effects. Wear also will change the dimensions, and is likely to result in irregular performance. Extreme precautions to keep grit out of the fluid are required. The dead weights may lose weight from wear, or may gain weight by oxidation or collection of dirt.
Height and Piston-Buoyancy Corrections
It usually happens that the gage being tested, or the point at which the pressure is to be measured, is not at the sae level as the lower end of the piston. Therefore correction should be made for the pressure difference due to the head of oil between these points. The correction is negative when the gage is above the piston. When oil is used in the piston gage, the correction will be approximately 0.03 psi for each inch difference in level.

When the submerged part of the piston is of uniform cross section, the pressure measured is that at the level of the lower end of the piston. In some designs the piston is enlarged to provide a stop for its upward motion or to give increased strength. If these enlargements are submerged in liquid, the weight of the fluid displaced by the enlargement should be subtracted from the dead-weight load.

With some designs it is not possible to observe the liquid level, and therefore not possible to determine the submerged volume. In such a case it is necessary to determine the buoyancy correction by test. It will usually be less than a pound per square inch.

Pressure Drops in Lines
Whenever possible, the piston gage should be connected into a leak-tight system, with tubing of the largest bore consistent with strength and safety, in order to avoid pressure drops in the lines. If this is impossible, the layout of tubing should be such that the piston gage is connected to the test vessel with a line through which the lease possible flow of fluid occurs.
Friction
If the piston and cylinder come into contact, or if any part o the dead weight touches the supporting frame, sliding friction will introduce uncertainties. A film of fluid between piston and cylinder is usually maintained by rotation or oscillation. With rotation there is the possibility of a corkscrew effect, an axial thrust produced by helical irregularities on the surface of the piston and cylinder. Observations should be repeated with the piston rotating in the opposite direction. If the hydraulic fluid is a good lubricant and has a suitable viscosity, a film can be maintained with either rotation or oscillation. A lubricating film can be maintained with a fluid of very low viscosity (even air) provided the piston is rotated above a critical speed. This speed depends on the fluid, the symmetry of loading, and the clearances, etc. It must be determined experimentally for each instrument.

Intercomparison Techniques

Since the percentage accuracy of the measurements of the diameter is better for a large piston than for a small one, the intercomparison of piston gages of differing ranges provides an opportunity of improving the accuracy of the latter. In addition, there is the opportunity of averaging out some of the effects of varying properties of the materials of the piston, and of picking up possible mistakes in measurements.

In an intercomparison by cross floating, two piston gages are connected together and to a common pressure generator, such as the intensifier, of suitable range and capacity. Either piston gage can be isolated by means of valves in the connecting tubing. A telescope with an eyepiece scale or the equivalent, is set up to observe the position of one of the pistons in its range of motion. At the selected pressure, the rate of fall of each piston is observed with the gage isolated. The two gages are connected together and one of the loads is adjusted by adding small weights until both pistons fall at approximately the same rate as when isolated. An exact balance is usually not obtained, but by observing rates of fall at various loads, an interpolation can be made to the load for which the rates of fall would be the same with the gages connected as when they are isolated.

By having one of the pistons fall repeatedly over the same interval, while the other falls through various parts of its range it is possible to observe small variations in diameter. By this method a commercial piston gage with a simple cylinder was ascertained to have an effective area constant over its length to within 1 part in 50,000. Since the diameter of this piston was about 0.16 inch, it must have been uniform to 1.6 microinch or better.

This particular observation happened to have been made at a pressure of about 1000 psi. When the same kind of comparison was attempted at 5000 psi, the rate of fall was so much greater that the comparison was quite unsatisfactory. The enormous change in sensitivity was caused by an increase in the bore of the cylinder estimated to be only about 8 microinches.

Calibration Against a Mercury Column

An alternative primary standard is the mercury column. Readings on it are also subject to a large number of corrections most of which are counterparts of those for the piston gage. The reader may refer to Glazebrook (8) or Meyers and Jessup (5) for a discussion of them. The latter authors set up a 150-psi mercury column, with an accuracy of 0.01 percent, compared it with piston gages of 0.01 percent accuracy, and had agreement within 0.01 percent at pressures up to 1050 psi. They set up a pressure on one piston gage, added the pressure measured by the mercury column, and compared the sum with the pressure on a second piston gage. The pressure measured by the mercury column was then added to that on the second gage and compared with a new pressure set up on the first. The process was continued until the glass tubing broke. Keyes (9) has carried a similar procedure up to pressures as high as 15,000 psi.

While this method is quite satisfactory at low pressures, the accuracy falls off rapidly at higher pressures. For example, suppose that a 150-psi mercury column is used to calibrate a pair of 30,000 psi piston gages, which have a sensitivity of 1 ppm of their range or 0.03 psi. Then each 150-psi step in this comparison can be done to 0.03 psi, or 1 part in 5000. But the area of a piston good enough to attain this sensitivity probably can be measured to better than 1 part in 10,000.

The small distortion of the cylinder or piston in the 150-psi step of pressure gives rise to just as large a correction, in relation to the 150 psi, as the larger distortion under 30,000 psi is with respect to the 30,000 psi range of the instrument.

A satisfactory comparison with a mercury column will give the user of the piston gage the valuable assurance that he has not overlooked some large systemic error in the latter. But a calibration of a piston gage against a mercury column cannot be regarded as inherently better than a standardization by direct measurement on piston and weights.

Fixed Points on the Pressure Scale

Primary pressure standards are not portable. In order to attain an overall accuracy of 1 part in 10,000, attention must be paid to a great many factors. As a result, although the mercury column or piston gage is simple in principle, it will be surrounded by so many auxiliaries that its bulk will be large in the aggregate. To tear it up by the roots and send it to another laboratory for a calibration is a formidable task. Even if all this were done, there would be no assurance that essential parts would endure the hazards of shipment. There is a real need for a means of calibration that can be put in a suitcase, or even better, sent by first class mail.

One solution to this problem is a set of experiments which can be repeated at any laboratory and will reproduce definite pressures. In some cases a suitable transfer instrument and a set of fixed-pressure points would replace the primary standard. The analogous situation exists in the field of temperature measurement. The platinum-resistance thermometer and four fixed-temperature points define the temperature scale over a range of 600°C to the satisfaction of most users.

The fixed-pressure points may be based on changes of state or polymorphic transformations which should satisfy the following conditions:

  1. The substances involved should be obtainable in a pure state.
  2. The transition should take place at a sharply defined pressure, with no range of indifference.
  3. There should be a large volume change, or other effect, for ready identification.
  4. The reaction should run rapidly.
  5. The dependence of the pressure on temperature or other possible parameters should be slight.
  6. It should be possible to contain the substance so as not to damage the pressure vessel.
Triple points are particularly suitable since they are unique in both temperature and pressure, provided all three phases are present. In a two-phase system a volume change in associated with a thermal change, and time is required to approach equilibrium. At a triple point, adjustments among all the components take care of thermal and volume changes simultaneously.

Bridgman's value of the freezing pressure of mercury at 0°C, 7640 kg/cm² (108,660 psi), has been used by a number of observers. Occasionally a higher temperature has been used when a higher pressure was required. The temperature coefficient of the melting pressure is about 3000 psi per degree C, so that accurate temperature measurements and precise control are required. These are much easier at 0°C, where an ice bath can be used.

In his work at the high pressures Bridgman has used the bismuth I-II transition at about 360,000 psi; L-III-V near 50,000 psi; L-V-VI near 90,000 psi, L-VI-VII near 320,000 psi. Between these points the melting-point pressure measurements on the basis of a value of 9630 bars (139,670 psi) for the freezing pressure of water at 30°C, with the expectation of adjustment if a better determination should change that value.

Professor Bridgman measured pressures at the phase changes of water and of mercury in 1912. Since then, in so far as is known, there has not been an independent determination. Rather, the value reported by him has been used without question. While the Bridgman values were sufficiently accurate for his pioneering work, a vast amount of data reported with greater accuracy than that claimed by Bridgman are still dependent on his value. This is not as serious as it might seem at first glance. Bridgman's value should stand until an independent determination has been made with the improved tools of the present day. Quantitative work reported prior to the publication of the new value will have to be corrected when correlated to subsequent work—should a correction for Bridgman's value be indicated.

Mercury-Point Determination

In order to test the possibilities of the controlled-clearance piston gage, and to get the "feel" of the experiment, a determination was made of the melting pressure of mercury at 0°C. Use was made of a transfer gage, at first a manganin-wire pressure cell, and for the final runs, a gold-chromium-wire pressure cell. The determination can be broken into three steps as follows:
  1. Measurement of melting pressure of mercury in terms of transfer instruments.
  2. Calibration of transfer instruments in terms of piston gage.
  3. Direct measurement of area of piston and mass of dead-weight load.
Mercury-Point Comparisons
The apparatus found to be convenient was arranged as shown schematically in Figure 7. Two laboratory-type intensifiers were used. The intensifiers were driven and charged by hand-operated 10,000 psi pumps. White gasoline was the high-pressure fluid used. Kerosene had been found unsuitable in an earlier experiment. The upper side of the low-pressure piston of the intensifier A (normally vented to the air), was filled with a liquid and connected to a sight glass to indicate the piston displacement, and hence the changes in volume. With a 1/8-inch-bore sight glass, the displacement of the column was 360 times that of the piston.

The mercury bomb was a simple high-pressure vessel with a cavity 1 inch diameter and approximately 3 inches long. Freshly distilled mercury was sealed in a polyethylene sack and placed in the bomb cavity.

A gold-chromium resistance element, subjected to the high pressure, was used as a transfer device. The gold-chromium resistance was part of a 120-ohm equal-arm bridge. The other arms were in a bath at room temperature, close to the pressure cell. The bridge was driven by alternating current at 1000 cps, 50 milliamp total current. One arm of the bridge was shunted by an adjustable resistance box. An a-c null indicator was used to detect balance. The sensitivity of the indicator was adequate to detect changes in the resistance of the gold-chromium wire of 1 part in 107. A duplicate bridge in a temperature-controlled bath was used to correct for drift in the null indicator. The drift during a run was not more than 1 part in 106.

The system was charged to about 10,000 psi by a hand pump. This pressure was held in the system by a check valve when the hand pump was removed. The piston of intensifier (B) was then advanced to the end of its stroke raising the pressure to about 90,000 psi. Another check valve held the pressure in the rest of the system after the pressure in intensifier (B) was released. During the rest of the experiment the pressure was raised by advancing the piston of intensifier (A). The movement of the piston of this intensifier was followed in the sight glasses.

If the volume is plotted against pressure, at constant temperature, there is obtained a nearly straight line A-B, Figure 8, representing the compression of the liquid, a horizontal straight line B-C, representing the change in volume at constant pressure of the mixture of liquid and solid, and the nearly straight line C-D representing the compression of the solid. This is under isothermal conditions, which are never realized by the experimenter. When the specimen is compressed, it is heated, and is therefore at a pressure which is a little high. If only one phase is present, either liquid or solid, the approach to equilibrium is nearly completed after 5 minutes. With both phases present, the latent heat of the phase change is large, and relatively long time is required for the system to come to equilibrium as this heat is transferred to the bath.

In these measurements the pressure is increased in steps of about 5000 psi, and the piston is held stationary at each step for 5 minutes. As long as only liquid mercury is present, the pressure drops back slightly and stabilizes during each hold period, A to B, Figure 8. When solid mercury makes its appearance, the pressure drops back farther in the 5-minute period and the curve drops off, as at E. When all-solid line C-D is reached, there is an abrupt transition to the quick equilibrium condition, and on further compression, the curve resembles A-B.

On expansion, a zig-zag line D-C-F-A would be described. If at the point G, the volume were held constant the pressure would settle out toward the equilibrium point. The time of several hours required for a close approach to equilibrium may exceed the experimenter's patience. A quicker way is to juggle pressures so as to start close to the expected asymptote, as at H or I. Then the equilibrium pressure can be approached from both sides, the decay being followed only long enough to predict the position of the asymptote. This should be repeated at several points along the line B-C, including one point close to C. If the mercury is impure, the corner at C will not be sharp, as in the dotted line.

In addition to the plot of pressure (or what is more convenient, readings on the transfer gage) against volume, it is also desirable to plot pressure against time. In Figure 9 are such curves. In (a) only liquid is present, and the decay is rapid. This is a leak check showing a loss of about 3 psi per minute after disappearance of the initial transient following an expansion; (b) was taken near (F) with both phases present; (c) and (d) were taken at (H) and (I) and define the melting pressure as their common asymptote.

It is desirable that the system be absolutely tight. If the leak in, say, 30 minutes cannot be detected by the pressure-measuring means when only one phase is present, its effect may be neglected. If it is perceptible but small, its effect can be estimated and a correction applied.

From a curve such as (a), Figure 9, with only one phase present, the leak in psi per minute may be estimated. The flow through the leak is supplied by the expansion of the fluids, mostly gasoline, and the contraction of the pressure vessel under this pressure change. The fall of pressure with both phases present, as (c) in Figure 9, can be divided into two parts, the pressure drop associated with the expansion of the fluid under the leak, and that associated with the change in volume due to the phase change. When the pressure is falling rapidly, as at the beginning of curve (c), the mercury is freezing, the sample is warmer than its surroundings and is giving up heat to the pressure vessel. When the pressure is falling very slowly, the fluids are expanding very slowly, and most of the leak is supplied by the expansion which results from melting of the sample. The sample then will be absorbing heat from the pressure vessel and will be a little cooler. At the point of (c) where the slope is the same as that of the leak curve, the sample is neither melting nor freezing and its temperature is the same as that of its surroundings. The difference between the pressure at this point and that at the asymptote is approximately the leak correction.

An alternative and equivalent estimate of the effect of leak is obtained by multiplying the rate of the leak from curve (a) by the time constant of the decay of pressure in curve (c) or (d).

Calibration of the Transfer Instrument
The arrangement of apparatus used in calibrating the gold-chromium pressure cell is shown in Figure 10. A pressure was generated by intensifier (A), and admitted to the controlled-clearance piston gage (B). This pressure was indicated on the gold-chromium pressure cell (C) with the same bridge and null indicator used in the observations on the mercury point. The jacketing cylinder of the piston gage was subjected to pressures generated by intensifier (D) and measured by a manganin cell (E) which was part of a Wheatstone bridge of the strain-indicator type.

A dead-weight load was set up on the piston gage, the pressure raised until the piston was at the top of its travel. While the piston fell through its range (1/8 inch) observations were made on the pressure under the piston with the gold-chromium cell, on the jacketing pressure with the manganin cell, and on the time of fall. The jacketing pressure was adjusted until the rate of fall was about 30 seconds. Readings were made at the top and the bottom of the travel. The dead-weight load on the piston was increased in 20-lb. steps until the pressure used in the mercury-poing measurements had been exceeded. The load corresponding to the mercury point was obtained by interpolation between calibration points on the basis of the resistance change in the gold-chromium cell.

Measurement of Piston and Dead Weights
The 20-lb. weights had been adjusted by the Massachusetts State Sealer of Weights and Measures to within 0.001 lb. of their nominal mass. The yoke, weight hanger, piston, and so on, were weighed to an accuracy of about 0.001 lb.

The piston was measured by the Gage Section of the National Bureau of Standards with an accuracy of about ±0.00002 inch.

Results
The results of the measurement and estimates of the various errors are summarized in Table 2. The errors characteristic of piston gages, in general, have already been discussed at length.
 
Table 2: Summary of Results
Weights: 268.73 lb.
Area: 0.0024446 square inches
Pressure, uncorrected: 109,930 psi
Correction, psi
Residual Error, psi
Zero shifts, in mercury-point runs
...
±65
Zero shifts, in piston-gage runs
...
±215
Correction for leakage
+20
±10
Piston measurement, ±0.00002 inch
...
±80
Piston taper, 0.00003 inch
...
±120
Residual clearance, 0.000010 to 0.000020 inch
-30
±10
Elastic distortion
+65
±10
Temperature of piston
-25
±15
Weight measurement
...
±10
Buoyant effect of air on weights
-15
 
Gravity at Foxboro, Massachusetts, USA
-35
±5
Temperature of ice bath, ±0.02°C
...
±60
Conduction down tubing
-150
±150
Total Correction
-170
 
Corrected Pressure
109760
±750
The mass load on the piston gage at the pressure for which the resistance change was the same at that observed in the mercury fixed-point runs was obtained by interpolation between the observations in the piston-gage runs. The area was the average in the upper ¼ inch of the piston, based on direct measurement. The uncorrected pressure is the mass load divided by the area.

There were differences in zero before and after the mercury-point and the piston-gage runs. The two zero readings were averaged, on the assumption of equal changes during increase and decrease.

The system was found to be losing pressure at a rate of about 2 psi per minute when only liquid was present. This rate was multiplied by the 10 second time constant of the approach to equilibrium conditions when both phases were present.

The residual clearance between piston and cylinder was estimated from the leakage rate, including the effect of the expansion of the fluid under the pressure change during the fall of the piston, and assuming that the drop in pressure occurred in 1/10 inch of length in the piston.

The elastic distortion of the steel piston was calculated from Equation 10.

The value of gravity was interpolated from Coast and Geodetic Survey data at points within 10 miles of Foxboro, Massachusetts, where the experiment was conducted.

The temperature of an ice bath, using good commercial ice, usually can be expected to be 0.00 ± 0.02°C. Several measurements of samples during these tests fell within this range. The conduction of heat down the pipe to the mercury cell would raise the temperature of the latter by an amount estimated not to exceed 0.1°C.

A difference between pressures at the upper and lower limits of travel of the piston was observed. This was found to be caused by a taper in the piston which developed when the piston had been scored in previous tests. (This taper of 0.00012 inch in ½ inch of length was 10 to 100 times that which would be tolerated in a new piston.) The ratio of 0.7 between change of jacket pressure and change of measured pressure indicated that the line of contact between piston and cylinder was near the top. The area of the piston and cylinder was near the top. The area of the piston was computed on the diameter of 0.05579 ± 0.00002 inch measured 1/8 inch from the top. An additional uncertainty in area was taken as that corresponding to the taper in 1/8 inch of length.

Additional possible errors, not evaluated, include those resulting from possible changes in the resistance in the gold-chromium wire which were not reflected in the zero shift. This alloy is a new one for pressure measurement, and very little experience has been accumulated. In particular, little or nothing is known as to the proper pressure-seasoning treatment. (This particular element had been subjected to two applications of pressure to 140,000 psi.)

Difficulties Encountered
A number of mechanical difficulties resulted from the escape of mercury from its container in an early experiment. A small quantity got into other parts of the system, and amalgamated with brass packing rings, the heads of pistons and the like. Numerous failures of these parts resulted. No autofrettaged cylinders failed. It was found that a polyethylene bag was tough and flexible enough to hold the mercury. Whenever possible, the inlets to pressure cells, intensifiers, and the like, should be at the bottom, and the inlet to the mercury bomb at the top, and the tubing arranged so that escaped mercury droplets will drain to the mercury bomb.

In the manganin and gold-chromium cells the packing used for the insulated electrical lead included a plastic which softened in contact with kerosene or gasoline, and was responsible for a progressively increasing electrical leakage. This was serious since one corner of the Wheatstone bridge is grounded, and a leakage through 100 megaohms will produce a significant error. Other packing materials are being tried.

Conclusions

The measurements of the melting pressure of mercury were of value in trying out new equipment in the high-pressure field and in pointing out the direction of future improvements in technique and apparatus.

The controlled-clearance piston gage was used at pressures as high as 120,000 psi, with no indication that the limit of its range was being approached. (The pressures to which this instrument was used were limited by fittings rated at 100,000 psi.) Adequate floating times for the measurements were obtained. Indications were that clearances could be held to a few microinches, provided the surfaces of the piston and cylinder were that good. The small clearances make necessary precautions to eliminate abrasive particles in the working fluid. The use of carboloy pistons may be indicated from considerations of wear.

The gold-chromium pressure cell seemed promising. The time required to reach constant reading seemed somewhat shorter than for manganin. The very small temperature coefficient should be valuable. The elastic defects of hysteresis, drift, and after-effect, required further study since they were obscured by the electrical leakages in other parts of the bridge.

These measurements give a value 1 percent higher than obtained by Bridgman for the melting pressure of mercury. This difference is not serious in view of the rather large experimental error of ±0.7 percent due to identifiable sources.

The largest errors can be reduced substantially in future work. A better choice of the insulating material in the electrical leads should reduce the troubles with zero shift. An unworn piston would have variations in diameter less than a tenth as great as those in the one used. The temperature of the mercury sample can be measured directly.

Acknowledgments

This work was carried out with the support of the Foxboro Company, Harwood Engineering, Inc., and the National Bureau of Standards. The authors are indebted to numerous people in those organizations for assistance. They particularly wish to thank Dr. Brombacher of the National Bureau of Standards, and Mr. Howe of the Foxboro Company for their encouragement; Mr. George Champagne for his technical assistance; Mr. Howard Fuller, who designed and operated the a-c null indicator used. The authors have profited from many suggestions made by Professor Bridgman.

Bibliography

1. "Mémoires sur l'élasticité et la dilatbilité des fluides jusqu'aux très hautes pressions," by E.H. Amagat, Annales de Chemie et de Physique, vol. 29, 1893, pp. 68-136 and 505-574.
2. "Einfluss der Rotation auf die Empfindlichkeit einer aboluten Druckwage," by A. Michels, Annalen der Physik, vol. 72, 1923, pp. 285-320; vol. 73, 1924, p. 577.
3. "The Measurement of Hydrostatic Pressures Up to 20,000 Kilograms per Square Centimeter," by P.W. Bridgman, Proceedings of the American Academy of Arts and Sciences, vol. 47, 1912, p. 321.
4. "Dead-Weight Machines of 111,000 and 10,100 Pound Capacities," by B.L. Wilson, D.R. Tate, and George Borkowski, NBS Circular C446, June, 1943.
5. "A Multiple Manometer and Pisong Gages for Precision Measurements," by C.H. Meyers and R.H. Jessup, National Bureau of Standards, Journal of Research, vol. 6, 1931, p. 1061, RP324.
6. "Units of Weight and Measure, Definitions and Tables of Equivalents," National Bureau of Standards Miscellaneous Publication, M121.
7. "Strength of Materials, Part II. Advanced Theory and Problems," by S. Timoshenko, The MacMillan Company, New York, NY, 1936.
8. Glazebrook-Dictionary of Applied Physics.
9. "Methods and Procedures Used in the M.I.T. Program of Investigation of the Pressures and Volumes of Water to 460°C," by F.G. Keyes, Proceedings of the American Academy of Arts and Sciences, vol. 68, 1933, p. 505.
10. "Equilibrium in Binary Systems Under Pressure I. An Experimental and Thermodynamic Investigation of the System, NaCl-H2O, at 25°," by L.H. Adams, Journal of the American Chemical Society, vol. 53, October, 1931, p. 3769.

Harwood Engineering Company, Incorporated – © July, 1999
455 South Street, Walpole, Massachusetts 02081
phone: 508-668-3600
fax: 508-660-2276
http://www.ultranet.com/~harwood/
email: harwood@ma.ultranet.com