Harwood
Engineering Company, Inc.
Calibration of Ballistic Pressure Transducers
by Charles D. Bullock and Arpad A. Juhasz
Current procedures used for the calibration of ballistic pressure transducers
at BRL are described. Checks include evaluation of continuity, hysteresis,
and zero return characteristics as well as calibration against a dead weight
system. Static versus dynamic response behavior is evaluated with the aid
of a high pressure dynamic positive step calibrator. For the most exacting
measurements, adapters are used permitting calibration of transducers in
the same mechanical environment as during measurement. Recommended recalibration
intervals are indicated.
Introduction
The mission of the Interior Ballistic Division of BRL includes research
on novel ballistic concepts, charge design methodology and advancing the
state of the art in interior ballistic computations. These efforts are
supported by a variety of combustion, interior ballistic and ballistic
simulator firings. Central to all these experiments is the measurement
of pressure. Pressures may range from a few hundred pounds per square inch
(psi) to a hundred thousand psi full scale depending on the experiment.
The quality of the measured pressures, in large scale, is dependent upon
the methodology, care and accuracy of the calibration process.
The primary function of the calibration procedures is to determine transducer
response characteristics and to act as a screening tool to help weed out
problem transducers before they can do damage. A secondary but vital function
is to help solve measurement related problems and assure that the devices
perform as required under the conditions of service. The purpose of this
paper is to discuss the procedures which have evolved over the past twenty
five years at BRL for the calibration, selection and use of high pressure
transducers for ballistic applications. It will include a discussion of
the most important characteristics of high pressure transducers, BRL's
calibration and evaluation procedures and a look at potential problem areas.
Discussion
Transducers Used
Ballistic pressure transducers in routine use at BRL fall into two categories,
piezoelectric element and single arm strain sensors. The commonly used
piezoelectric transducers (gages) are obtained commercially. The strain
sensor transducers are made privately for BRL. All of the above are used
daily to measure pressures up to 100,000 psi. The gages have a fast response
(10-90 percent response times on the order of 10 microseconds). The events
measured range from the sub-millisecond to several hundred millisecond
time frames.
High pressure transducers can, with adequate care in calibration and
use, be successfully employed to make measurements under 1000 psi. This
requires special calibration procedures, however, which will be discussed
later. In addition to high pressure transducers, good low pressure, fast
response transducers are also commercially available and find applications
in ignition simulators and the like. At the other end of the spectrum,
a current development effort is aimed at providing a ballistic pressure
transducer capable of measuring pressures to 200,000 psi.
Calibration Procedures
The purpose of pressure calibration is to determine the response of the
transducers to known pressures, to verify the response specified by the
manufacturer, and to show repeatability. During the calibration procedure
for a given transducer the following questions are considered:
-
is the response continuous
-
does it suffer from hysteresis
-
does it return to zero
-
is response linear or at least well behaved
-
is there a difference between first and subsequent cycles
-
are static and dynamic characteristics the same
-
have response characteristics changed with use
Initial Screening
Calibration typically begins with an examination of the continuity, hysteresis
and zero return properties of a transducer over the pressure range intended
for use. A schematic of the main calibration system in use at BRL is given
in Figure 1. The output of the test transducer
is plotted (Y-axis) against the output of a stable reference strain gage
transducer of known characteristics (X-axis) while the system is pressurized
and depressurized over the desired pressure range. Pressurization is accomplished
using the air pump/intensifier portions of the system. The response curve
of the transducer is used as an indicator of its overall quality.
Continuity
Examples of "good" and "bad" continuity response are given in Figure
2. In this case, both plots were obtained from the same transducer
but at different times, indicating degradation in performance as a function
of use. Normally, when discontinuities of this type are encountered, the
transducer is retired.
Hysteresis
Examples of "good" and "bad" hysteresis characteristics are shown in Figure
3. In the plot on the left the ascending and descending portions of
the curve coincide. In the plot on the right the transducer appears to
take a "set" on depressurization. Normally, a maximum hysteresis level
of 1-2 percent of full scale is thought to be acceptable. Excessive hysteresis
would make interpretation of the up and down slope portions of ballistic
data difficult to interpret.
Zero Return
Examples of "good" and "poor" zero return properties of a transducer are
given in Figure 4. In this case, the transducer
exhibits a residual output after the pressure loading is removed. Normally,
the maximum error in zero return deemed acceptable is one percent of full
scale.
A final characterization made at this point involves a comparison of
the first and subsequent cycles of transducer output. With certain transducers,
response characteristics change between the first and subsequent pressurization
for a given installation. This could lead to serious problems in measurements,
especially in cases involving cyclic events such as multi-shot bursts.
Examples of "good" and "bad" first vs. second cycle output are given in
Figure 5. Normally, a difference of less than one
percent full scale variation of cyclic output is found to be acceptable.
Quantitative Procedures
Quantitative transducer response characteristics are obtained using a dead
weight system. A simplified schematic of a dead weight system is given
in Figure 6. The principle of operation is simple.
At the point of equilibrium, that is, where the mass/piston combination
is exactly balanced by pressure in the fluid, the hydrostatic pressure
throughout the fluid may be calculated by dividing the total mass ("weight"
plus piston) by the piston area. The output of the test transducer is measured
at a series of float points corresponding to various mass loadings. The
static dead weight calibration method has both advantages and disadvantages;
it is the most accurate and repeatable source of calibration pressures.
Its principal disadvantage for ballistic applications is that it needs
dynamic verification.
The dead weight calibrator used at BRL is a 100 Kpsi Astra model D100KS.1
In this device the pressurizing fluid balances the force of a series of
calibrated masses transmitted through a piston of precisely known area.
A thin film of hydraulic fluid separates the piston from the cylinder wall
and the piston is oscillated about its axis to reduce the effects of static
friction. Various combinations of masses permit the generation of pressures
at intervals as small as 100 psi up to a maximum pressure of 100,000 psi.
In the "at rest" position, all the masses are loaded onto the yoke. A series
of air operated lifters is used to download respective masses to yield
the desired pressures. The tare pressure (due to the weight of the yoke
alone) is three thousand psi, representing the minimum pressure attainable
with this system. Common calibration intervals are 5 Kpsi and above.
For the most accurate work with deadweight calibrators, corrections
need to be made for the effects of gravity, buoyancy of air, and effective
piston area of the calibration system.2 The maximum error from
all these sources for our facility is less than 0.25 percent, as measured
by cross-floating our device against a Harwood controlled clearance deadweight
calibrator. The precision of the Astra gage as used is 0.15 percent. Our
data analysis program is currently being revised to include the buoyancy
and gravity corrections. Once this is complete, the accuracy of the system
is expected to move closer to its precision.
The dead weight device as used in our main calibration station is given
in Figure 1. The pressure generating source is
an hydraulic intensifier (pressure multiplier) of 16:1 ratio, capable of
producing over 100 Kpsi. The low pressure side is driven by a 10 Kpsi air
pump. The normal high pressure medium is plexol 201 (now Monoplex), a synthetic
lubricant. A check valve, relief valve, additional valving and hydraulic
reservoir complete the system.
Transducers are calibrated over the range at which they will be used.
Typically, if the expected maximum pressure for an experiment is 100 Kpsi,
a series of points at 20 Kpsi intervals is chosen for calibration. For
a test series with an expected maximum pressure of 25 Kpsi, 5 Kpsi intervals
are used. Output values from both the upward and downward portions of the
calibration cycle are included, giving typically 11 points for curve fitting
purposes. In Table 1 the values Y1, Y2 and Y3 are voltage reading representing
transducer output which are averaged (Y) and converted to gage output units
(PCB), in this case picocoulombs. The data are fitted via a least squares
method to a first degree equation. Curve fitting is done both in terms
of transducer response vs. pressure and pressure vs. transducer response
(which is used in computerized data reduction programs). Due to the slight
curvature in even "good" pressure transducers, users generally prefer the
second order fit for computerized data analysis purposes. For the sake
of simplicity, however, they prefer the first order fits to make the amplifier
settings.
Table 1. Gage Calibration Record for Piezoelectric Gage
|
PT
|
KPSI
|
Y1
|
Y2
|
Y3
|
Y
|
PCB
|
|
1
|
0
|
0.0000
|
0.0000
|
0.0000
|
0.0000
|
0.0
|
|
2
|
20
|
0.2673
|
0.2673
|
0.2673
|
0.2673
|
2673.0
|
|
3
|
40
|
0.5399
|
0.5399
|
0.5399
|
0.5399
|
5399.0
|
|
4
|
60
|
0.8183
|
0.8183
|
0.8183
|
0.8183
|
8183.0
|
|
5
|
80
|
1.1069
|
1.1069
|
1.1069
|
1.1069
|
11069.0
|
|
6
|
100
|
1.4010
|
1.4010
|
1.4010
|
1.4010
|
14010.0
|
|
7
|
80
|
1.1090
|
1.1090
|
1.1090
|
1.1090
|
11090.0
|
|
8
|
60
|
0.8219
|
0.8219
|
0.8219
|
0.8219
|
8219.0
|
|
9
|
40
|
0.5435
|
0.5435
|
0.5435
|
0.5435
|
5435.0
|
|
10
|
30
|
0.2700
|
0.2700
|
0.2700
|
0.2700
|
2700.0
|
|
11
|
0
|
0.0006
|
0.0006
|
0.006
|
0.0006
|
6.0
|
|
First Degree Fits
|
|
PCB =
|
8.6081E+01
|
1.3946E-01
|
* PSI
|
|
|
|
|
PSI =
|
-6.3059E+02
|
7.1683E+00
|
* PCB
|
|
|
|
|
MPA =
|
-4.3489E+00
|
4.9436E-02
|
* PCB
|
|
|
|
|
Correlation Coefficient = 0.99985
|
|
Second Degree Fit
|
|
PSI =
|
-4.6285E+02
|
7.5573E+00
|
* PCB
|
-2.9739E-05
|
* PCB
|
2
|
|
MPA =
|
-3.1921E-01
|
5.2119E-02
|
* PCB
|
-2.0510E-07
|
* PCB
|
2
|
|
Correlation Coefficient = 0.99999
|
When high pressure transducers are to be used in low pressure measurements,
special precautions are needed. Certain makes and models of high pressure
transducers appear to have superior low pressure linearity and torque sensitivity
properties. These transducers are first prescreened using the conventional
technique. The best of these units are calibrated against a 10 Kpsi dead
weight system using the negative going pressure step method. That is, the
deadweight system is floated at a given pressure against the gage and the
pressure is released to zero. (The signal generated is equal to but opposite
in sign to the output during actual pressure measurement). The amplifier
is zeroed just prior to the pressure step, the whole process taking approximately
one second. This fast procedure helps to reduce the effect of drift which
can be a significant problem in using high pressure transducers at their
low end. For calibrations under 0.15 Kpsi a commercial air operated dynamic
calibrator is used.
Problem areas and solutions
Among the practical transducer problems of interest to ballisticians are
changes in response characteristics as a function of use, poor dynamic
performance, response changes with calibration range, difficulties involving
concentricity and depth tolerances in gage ports, and differential pressure
measurements. The purpose of this section is to highlight some of these
areas and to point out procedures which may help to prevent difficulties
before they occur.
Change of response characteristics with use.
It is normal for transducer sensitivity to change with use. Typically,
gage response decreases, and linearity and hysteresis properties may be
adversely affected. Presumably this is due to a gradual degradation of
the sensing element. This need not be a problem, however. Regularly scheduled
recalibration is used to keep track of these changes, users changing calibration
constants as appropriate. For high pressure firings (90-100 Kpsi) recalibrations
are recommended at 5-10 round intervals. At lower pressures, say 60 Kpsi,
recalibration after 25-50 rounds is recommended. In extreme cases, degradation
can be sufficient to affect the continuity of response of the transducer
(see Figure 2). In such cases the device is immediately
retired. In other cases gages are retained until they fall outside of acceptable
hysteresis or zero return characteristics.
Dynamic Performance.
Questions of the dynamic vs. static response behavior of ballistic pressure
transducers have concerned ballisticians for some time. In an effort to
address this problem, BRL in conjunction with the Harwood Engineering Company
has developed a 150 Kpsi positive step calibrator.3 The device
has been used to assess transducer dynamic performance properties. In most
cases static and dynamic properties have agreed quite well. Occasionally,
however, problems have occurred. Figure 7 illustrates
examples of "good" and "bad" dynamic response behavior. In the bad response
case the signal from the transducer is showing an upward creep over many
milliseconds. Similar, but shorter term, instances have also been observed
where the 10-90 percent response appear to be normal but the last 10 percent
of the response curve takes several milliseconds. It is thought that seal
movement and air bubbles under the strain patch may have caused some of
these problems. Other cases where dynamic response problems have occurred
have involved eccentric loading of gages due to mismatch between transducer
and mounting cavity, see below.
Calibration over the Wrong Range
A common error among project engineers is to use calibration data obtained
over one range in analyzing the results of experiments over a significantly
different pressure range. Table 2 illustrates this point. The gage in question
was calibrated to 100,000 psi and the calibration data fitted to a first
and a second order equation. The error columns indicate the the difference
between the fitted equation and the measured calibration points. The greatest
derivation from the curves comes at the low pressure end. Use of this calibration
data for measurements made over the 0-20 Kpsi range, for instance, could
introduce as much as six to seven percent error in the interpreted data.
This error can be dramatically reduced by recalibrating the transducer
over the 0-20 Kpsi range, see Table 3.
Table 2. Examples of Potential Errors Introduced at the Low Pressure
and in Calibration Over a Wide Pressure Range
|
PT
|
PSI
|
PCB
|
1ST
|
ERROR %
|
2ND
|
ERROR %
|
|
2
|
20000
|
2673.000
|
19160.83
|
4.20
|
19988.22
|
0.06
|
|
3
|
40000
|
5399.000
|
38701.59
|
3.25
|
39935.09
|
0.16
|
|
4
|
60000
|
8183.000
|
58658.10
|
2.24
|
59850.15
|
0.25
|
|
5
|
80000
|
11069.000
|
79345.78
|
0.82
|
80008.23
|
0.01
|
|
6
|
100000
|
14010.000
|
100427.72
|
0.43
|
100040.83
|
0.04
|
|
7
|
80000
|
11090.000
|
79496.32
|
0.63
|
80153.10
|
0.19
|
|
8
|
60000
|
8219.000
|
58916.16
|
1.81
|
60104.66
|
0.17
|
|
9
|
40000
|
5435.000
|
38959.65
|
2.60
|
40195.55
|
0.49
|
|
10
|
20000
|
2700.000
|
19354.38
|
3.23
|
20187.96
|
0.94
|
Table 3. Examples of Improvement in Error Levels Due to Calibration
Over the Range of Test Pressure
|
PT
|
PSI
|
PCB
|
1ST
|
ERROR %
|
2ND
|
ERROR %
|
|
2
|
5000
|
651.000
|
4902.81
|
1.94
|
4969.78
|
0.60
|
|
3
|
10000
|
1311.000
|
9873.40
|
1.27
|
9979.62
|
0.20
|
|
4
|
15000
|
1975.000
|
14874.11
|
0.84
|
14990.72
|
0.06
|
|
5
|
20000
|
2642.000
|
19897.42
|
0.51
|
1995.06
|
0.02
|
|
6
|
25000
|
3313.000
|
24950.86
|
0.20
|
24999.70
|
0.00
|
|
7
|
30000
|
3990.000
|
30049.48
|
0.16
|
30018.87
|
0.06
|
|
8
|
25000
|
3320.000
|
25003.57
|
0.01
|
25051.75
|
0.21
|
|
9
|
20000
|
2651.000
|
19965.20
|
0.17
|
20062.39
|
0.31
|
|
10
|
15000
|
1986.000
|
14956.96
|
0.29
|
15073.49
|
0.49
|
|
11
|
10000
|
1322.000
|
9956.24
|
0.44
|
10062.87
|
0.63
|
|
12
|
5000
|
662.000
|
4985.65
|
0.29
|
5053.51
|
1.07
|
Mounting Problems
Figure 8 and Figure 9 present
two common problems in transducer/mounting cavity interactions. A lack
of concentricity in either transducer or the mounting hole can result in
eccentric loading. The curves may look "normal" but the response may be
far out of line. Similar results can be noted if the sensing element touches
the bottom of the mounting cavity. In the case of one popular commercial
transducer, for instance, seal rings of two thickness, steel (0.010 inch)
or copper (0.020 inch), are available. The cavities are dimensioned according
to the intended seal thickness. Substitution of the thinner steel seal
for the originally intended copper, for instance, can cause the transducer
to "bottom" giving false, often high, readings. While this may seem, on
the surface, trivial, such problems are often hard to track down in practice
due to the depth and inaccessibility of gage cavities. A related problem
where tolerances are close is the flexing of the fixture in such a way
as to introduce transient mechanical loading on the transducer. Some mounting
problems may indicate their presence during installation. If the "feel"
of the transducer being screwed into the cavity is "too tight" chances
are that there is a concentricity problem. Alternatively, recording electronics
may be connected to the gage prior to installation to see if excessive
signals are generated during the mounting procedure.
Differential Pressure Measurements
One of the most exacting pressure measurement problems in gun ballistics
involves so called differential pressure measurements. Typically, the objective
is to measure pressure differences between the fore and aft ends of the
breech section in the early portion of the ballistic cycle to detect the
formation of pressure waves which may have ill effects on gun performance.
The problem is that whereas the event may have a maximum pressure of 80-90
Kpsi, the region critical to pressure wave formation is often below 10
Kpsi. The pressure differencing needs to be accurate at the low end of
the range, where gage errors are greatest. For these measurements, transducers
are preselected for the best linearity and hysteresis characteristics.
In addition, two and sometimes three sets of calibrations data are used
for the same transducer depending on the pressure range are used for the
same transducer depending on the pressure range being probed. For instance,
calibration data for 10 Kpsi maximum may be used to interpret the low pressure
end of the data while calibration data for 100 Kpsi maximum are used to
interpret the overall character of the full pressure-time curve. An additional
technique used to obtain quality pressure difference data involves the
use of mounting adapters, see below.
Mounting Adapters
One effective way of minimizing mounting problems is by use of an adapter.
A typical example appears in Figure 10. The small
size of the adapter permits easier machining and quality checks of the
cavity dimensions. A further advantage is that calibration can be done
in the adapter, making transducer remounting unnecessary. The adapter,
in effect, protects the transducer from mounting strains in the fixture.
If necessary, the measurement system, gage and amplifier can be calibrated
together, further refining the data. This can result in excellent performance
and repeatability. Figure 11 shows the response
behavior of a transducer in its adapter over three calibration cycles.
The lines are indistinguishable. The impressive fact about this data is
that the transducer had been used to make measurements between the calibration
cycles.
Summary and Conclusion
The procedures in use at BRL are aimed at preventing problem transducers
from entering the system and weeding out those whose useful life is past.
Continuity, linearity, hysteresis, and repeatability characteristics are
evaluated for new transducers and for transducers submitted for recalibration.
Numerical data are derived using a dead weight calibration system. Static
vs. dynamic performance differences, when suspected, are evaluated using
a high pressure, dynamic positive step calibration system. Calibrations
are recommended at 5-10 round intervals when pressure in the range of 90-100
Kpsi are to be measured. For pressures in the 60 Kpsi regime recalibration
is recommended at 25-50 round intervals. Calibration of transducers for
the expected pressure range is strongly recommended, since needless large
errors may be introduced at the low pressure end by the fitted curve. Mounting
adapters not only help to eliminate undesirable mounting effects on the
measurements but permit calibration of the transducer in the mechanical
environment of the actual measurement. This procedure has been especially
useful in exacting measurement applications such as differential pressure
measurements.
Acknowledgments
We wish to thank Mr. D. Dykstra, mentor t one of us (CDB), who was responsible
for setting up the original calibration system. Further thanks go to Harwood
Engineering Company, for their assistance with the controlled clearance
measurements and the dynamic positive step calibrator. Final appreciation
goes to R. Tompkins and J. Newberry for their assistance with this report.
References
1. Thirteenth Transducer Workshop, Monterey, CA
2. The Piston Gage as a Precise Pressure Measuring
Instrument; Harwood Engineering Company, Walpole, MA
3. Astron Gage Manual; Pressure Products Industries, Warminster, PA
4. Handbook of Transducers for Electronic Measuring Systems; Harz No.
Norton, Prentice Hall Inc., Englewood Cliffs, NJ
Harwood Engineering Company, Incorporated – © July, 1999
455 South Street, Walpole, Massachusetts 02081
phone: 508-668-3600
fax: 508-660-2276
http://www.ultranet.com/~harwood/
email: harwood@ma.ultranet.com